Method for the determination of benzene metabolite t,t-Muconic acid in urine by HPLC-UV with an Ion exclusion column.
dc.contributor.author | Gomes, Rafaela de Paiva | |
dc.contributor.author | Sanson, Ananda Lima | |
dc.contributor.author | Lobo, Fabiana Aparecida | |
dc.contributor.author | Afonso, Robson José de Cássia Franco | |
dc.contributor.author | Coutrim, Maurício Xavier | |
dc.date.accessioned | 2018-03-13T14:54:00Z | |
dc.date.available | 2018-03-13T14:54:00Z | |
dc.date.issued | 2016 | |
dc.description.abstract | The benzene metabolite trans,trans-muconic acid (tt-MA) is widely used as a biological indicator of exposure to this xenobiotic. An analytical method was developed and validated for the determination of urinary tt-MA using solid phase extraction onto a Strata SAX cartridge. The analysis was performed by HPLC with an Aminex HPX-87H ion exclusion column and a UV diode array detector. The sample preparation conditions were optimized using a 24 1 factorial design. Comparison of the slopes of standard analytical curves prepared in aqueous solution and urine showed that the biological matrix suppressed the tt-MA signal by around 50%, so the analytical curves were prepared with tt-MA standard at low concentrations in pooled urine. The analytical curves in the range of 5–500 g L 1 showed determination coefficients values (R2) > 0.99 for tt-MA standard in water and R2 > 0.98 for tt-MA standards in pooled urine. The coefficients of variation obtained using seven replicates were lower than 3.6%, and recoveries of tt-MA from solutions containing 5, 25, 50 and 100 g L 1 of the analyte were in the range 85%–90%, demonstrating the satisfactory precision and accuracy of the method. The limits of detection and quantification were 0.11 and 0.36 g L 1, respectively. The benefits of this new method developed is the possibility of complete chromatographic peak separation for the determination of tt-MA at baseline, without matrix components’ interference as normally found in the C18 column. This is the first time that this chromatographic column has been used for the analysis of tt-MA in urine. | pt_BR |
dc.identifier.citation | GOMES, R. de P. et al. Method for the determination of benzene metabolite t,t-Muconic acid in urine by HPLC-UV with an Ion exclusion column. Separations, v. 3, n. 2, p. 14, 2016. Disponível em: <http://www.mdpi.com/2297-8739/3/2/14>. Acesso em: 15 set. 2017. | pt_BR |
dc.identifier.doi | https://doi.org/10.3390/separations3020014 | |
dc.identifier.issn | 2297-8739 | |
dc.identifier.uri | http://www.repositorio.ufop.br/handle/123456789/9607 | |
dc.language.iso | en_US | pt_BR |
dc.rights | aberto | pt_BR |
dc.rights.license | This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/). Fonte: o próprio artigo. | pt_BR |
dc.subject | Exposure biomarker | pt_BR |
dc.subject | Environmental exposure | pt_BR |
dc.subject | Benzene | pt_BR |
dc.title | Method for the determination of benzene metabolite t,t-Muconic acid in urine by HPLC-UV with an Ion exclusion column. | pt_BR |